Modern easy-to-use expert class ECG machine with notebook design. Designed for both in-hospital and mobile use. 12.1" high-resolution color touchscreen foldable display. Provisional diagnosis output. Automatic data interpretation.



Preview and print
It is  possible to preview a report for both real-time printing and file storage.


Waveform process
Doctor can set the printing style, printing speed, sensitivity and use filter to get better waveforms instead of repeating the examination.


Replay and record
Stop and review the waveform backward and forward per second or page then print out to grasp every arrhythmia.


File storage
Internal storage for up to 500 patient records including waveforms, measurement and diagnosis, which can also be enlarged by flash disk or PC hard disk.

 

   
  • 12.1" foldable antiglare color touch screen
  • 12-channel ECG waves are displayed and recorded simultaneously
  • Curves "pause" and "repeat" for arrhythmia episode determination
  • Internal storage for up to 500 ECGs
  • Waveforms and auto-diagnosis results preview before printing
  • Heart rate variability (HRV) analysis
  • Data transmission to PC via series port Ethernet/USB/RS232/internal Wi-Fi
  • Flexible Printing Solution (built-in high resolution thermal recorder to print on normal thermal paper and A4 size fax paper+ support external USB printer)
  • Anti-noise technology with comprehensive filters and the high CMRR 121 dB
  • Stress test (Optional)
  • Trolley and cable bracket
  • ECG preview before printing

 

 

Specification:

 

Lead

12 Standard leads

Input Circuit

Floating input circuit, protection from defibrillation and pace-maker

Acquisition mode

12 leads simultaneously

Frequency Response

0,05Hs – 150 Hz

Input Impedance

>50MΩ

Input Circuit Current

10 nA

Noise level

<15μVp-p

Anti Baseline Drift

Automatic

Filter

EMG-Filter; 25Hz /35Hz /45Hz /Off

DFT-Filter 0,05/0,15/0,25/0,5Hz

Lowpass filter 150Hz/100Hz/75Hz

AC-Filter 50Hz/60Hz

CMRR

 >121 dB                                      

Recording Mode

Auto/Manual

Safety Standards

IEC 60601-1:2005, EN 60601-1:2006, EN 60601-1-2:2007
IEC 60601-1-2:2007, IEC/EN60601-2-25

Power supply

AC: 220 -230V, 50-60 Hz
Built-in rechargeable Li-ion battery: voltage 14.4 V

Record Paper

210mm x 30m, Z-Typ, in rolls                    

Paper speed

5;6,25;10;12,5;25;50mm/c

Dimension

420x 330x 105 mm

Display

800 x 600 pixels LCD-interface

Time constant

>3,2 s

Communication Interface

  Ethernet, RS 232, USB                   

Lead for rhythm test

1 or 3 lead-Option

Weight

6.5 kg

 

 

 

 

 

 

Basic service