This electrically powered and electronically controlled ventilator is designed for use in the ICU.

Features:

  • Suitable for children and adults
  • 8 inch color screen
  • Three level light and sound alarm
  • Sensitive pressure and flow trigger
  • Built-in battery

 

 

Parameter

 

Ventilation modes

Volume Controlled (VCV), Assist/Control

Pressure Controlled (PCV), Assist/Control

Synchronized Intermittent Mandatory Ventilation (SIMV)

Pressure Support (PSV)

Spontaneous Ventilation (SPONT)

Continuous Positive Airway Pressure (CPAP)

Combined: VCV+SIGH, SIMV (VCV)+PSV

Tidal volume

50-1500 ml

Respiratory rate

1-70 bmp (VCV, PCV), 1-40 /min (SIMV)

Inspiration time

0.2-6 s

Inspiratory pause time

0-2 s

FiO2

21-100%

Trigger sensitivity (flow)

2-30 l/min

Trigger sensitivity (pressure)

-2-0 kPa (above PEEP)

PEEP

0-3 kPa

PSV

0-6 kPa

PCV

0.5-6 kPa

Compliance

4ml/100Pa

Parameters monitored

Pressure values: Ppeak, Pplat, Pmean, PEEP

Volume/Flow values: Vti,VTe, MV, MVspont

Time values: ftotal, fspont, I:E, RSBI (option)

Oxygen Concentration: 21-100%

Real time curves: Pressure/time, flow/time, volume/time

Alarm

MV high/low

Leakage

Paw high/low

Airway pressure constantly too high

VT high

Frequency of spontaneous breathing too high

Apnea

Power failure

Battery low/exhausted

Gas supply failure

Gas supply

(O2) 0.28-0.6 MPa

Power supply

AC110~240V, 50/60 Hz, 65VA

Output Pressure

8 kPa

Noise level

≤65 dB

Interface

RS232

Dimensions (HxBxT)

350x303x280 mm

Weight

15kg

 

 

 