Infant incubator BabyGuard I-1103 for intensive care is air-temp and skin-temp servo-controlled by a micro-computer and equipped with a special air circulation system inside the apparatus which minimizes the flow in the hood environment. The automatic curtain reduces the heat loss. The double-walled hood conserves the heat for 2 hours.
Features:
- Mode of control: Air mode and Baby mode controlled by micro-computer
- Passive air humidification in the chamber
- Double-walled hood
- Basic stand on wheels and with brakes
- Weigh scales (optional)
- 3 drawers
- Side windows of the chamber are equipped with transparent and flexible sleeves.
- Front access window with iris diaphragm
- Smooth bassinet adjustment
- Film touchpad control
- LED indicators reflect real temperature of skin and air, set temperature of air, heating
- 7-alarm option
- RS-232 interface for distance control
Temperature servo-control | |
Air temperature control range | 25 – 37 °С (override mode 37 – 39° C) |
Skin temperature control range | 34 – 37 °C (temperature override mode to 37 – 38 °C) |
Temperature rise time | 30 min |
Inclination of the bassinet | ± 12° |
Airflow rate within hood environment | < 0,3 m/sec |
Noise level (inside hood) | ≤ 45 dB |
Alarms | Power failure Airflow failure Air temperature sensor failure Skin temperature sensor failure Temperature sensor deviation High temperature sensor failure Over temperature alarm |
Power supply | 220 – 230 V, 50 – 60 Hz |
Dimension | 1085mm ×640mm ×910mm (Main body)
1125mm ×665mm ×725mm (Cabinet)
|
Gross weight | 63kg
68 kg(Cabinet) |