Infant incubator BabyGuard I-1107 for intensive care is air-temp, skin-temp, air humidity and oxygen concentration (optional) servo-controlled by a micro-computer and equipped with a special air circulation system inside the apparatus which minimizes the flow in the hood environment. The automatic curtain reduces the heat loss. The double-walled hood conserves the heat for 2 hours.

Features:

  • Air temperature and skin temperature servo-control
  • Air humidity servo-control
  • Oxygen concentration display system
  • Double-walled hood
  • Basic stand on wheels and with brakes
  • Weigh scales (optional)
  • 3 drawers
  • Side windows of the chamber are equipped with transparent leaves and flexible sleeves.
  • Front access window with iris diaphragm
  • Smooth couch adjustment
  • Film control keyboard
  • LED indicators reflect real temperature of skin and air, set temperature of air, heating
  • 9-alarm option
  • RS-232 interface for distance control

Temperature servo-control

 

Air temperature control range

25 - 37 °C (temperature override mode to 37 - 39 °C)

Skin temperature control range

34 - 37 °C (temperature override mode to 37 - 38 °C)

Temperature rise time

30 min

Humidity control range

0 - 90 %

Indicator range of oxygen concentration

0 - 99 %

Inclination of the bassinet

± 12°

Airflow rate within hood environment

< 0,3 m/s

Noise level (inside the hood)

≤ 45 dB

Alarms

Power failure 
Airflow failure
Air temperature sensor failure
Skin temperature sensor failure
Temperature sensor deviation
High temperature sensor failure
High or low skin temperature
High or low humidity level
High or low oxygen concentration level

Power supply

220 - 230 V, 50 - 60 Hz

Dimension

1085mm×640mm×915mm (Main body)

1125mm×665mm×730mm (Cabinet)

Gross weight

68kg

60kg (Cabinet)

                                                 